HF-M6-3L

Closed-loop stepper motor drive basic industrial peristaltic pump, speed range 0.1-600rpm, flow rate 0.0053-3100ml/min. 8 different external control ways with RS232/RS485 communication interface (MODBUS protocal RTU mode). The pump can be connected with computer, PLC and HMI, achieve esternal control. Suitable for industrial sites, large flow rate, wide range and high precision liquid transferring. Mechanical keypad control start/stop, direction, timing test and external control selection. Rotary encoded knob control the speed or fast filling, easy to operate.

Characteristic





Industry application

Liquid transmission

Industry

Mineral processing supporting

OSD

Pump Head and Flow Rate

Pump Head Single/Dual Channel Tubing Wall Thickness (mm) Pressure Adjustable
HF-EasyPump III Single Channel 1.6 Adjustable
HF-EasyPump IV Single Channel 2.4 Adjustable
HF-EasyPump VI Dual Channel 1.6 Adjustable
Pump Head Single/Dual Channel Tubing Wall Thickness (mm) Pressure Adjustable
HF-EasyPump I Single Channel 1.6 No
HF-EasyPump II Single Channel 2.4 No
HF-EasyPump V Dual Channel 1.6 No
Pump Head Tubing IDxWall Thickness (mm) ml/r Flow Rate (ml/min) (0.1-600rpm) Tubing Max.Pressure (Mpa) Intermittent Tubing Max.Pressure (Mpa) Continuous Weight (kg)
Single Channel HF-EasyPump I/III 13# 0.8×1.6 0.053 0.053~32 0.27 0.17 0.6
Single Channel HF-EasyPump I/III 14# 1.6×1.6 0.27 0.027~162 0.27 0.17 0.6
Single Channel HF-EasyPump I/III 19# 2.4×1.6 0.55 0.055~330 0.27 0.17 0.6
Single Channel HF-EasyPump I/III 16# 3.1×1.6 0.933 0.093~560 0.27 0.17 0.6
Single Channel HF-EasyPump I/III 25# 4.8×1.6 1.967 0.197~1180 0.24 0.14 0.6
Single Channel HF-EasyPump I/III 17# 6.4×1.6 3.333 0.333~2000 0.14 0.10 0.6
Single Channel HF-EasyPump I/III 18# 7.9×1.6 4.3 0.430~2580 0.10 0.07 0.6
Single Channel HF-EasyPump II/IV 15# 4.8×2.4 1.8 0.180~1080 0.27 0.17 0.6
Single Channel HF-EasyPump II/IV 24# 6.4×2.4 2.733 0.273~1640 0.27 0.17 0.6
Single Channel HF-EasyPump II/IV 35# 7.9×2.4 3.833 0.383~2300 0.24 0.14 0.6
Single Channel HF-EasyPump II/IV 36# 9.6×2.4 5.167 0.517~3100 0.24 0.14 0.6
Single Channel HF-EasyPump V/VI 13# 0.8×1.6 0.0053 0.0053~32 0.27 0.17 0.6
Single Channel HF-EasyPump V/VI 14# 1.6×1.6 0.27 0.027~162 0.27 0.17 0.6
Single Channel HF-EasyPump V/VI 19# 2.4×1.6 0.55 0.055~330 0.27 0.17 0.6
Single Channel HF-EasyPump V/VI 16# 3.1×1.6 0.933 0.093~560 0.27 0.17 0.6
Single Channel HF-EasyPump V/VI 25# 4.8×1.6 1.967 0.197~1180 0.24 0.14 0.6

Parameter

Flow rate range 0.0053~3100 ml/min
Speed range 0.1-600rpm
Speed resolution 0.1 rpm, 0.1rpm; 100-600rpm, 1rpm
Flow rate accuracy <±0.5%
Back suction angle 0-360º
Display 3 digital LED
Motor type Closed-loop stepper motor
Control method Rotary encoded switch and Mechanical Keypad
Keypad lifetime 300.000 times
External control signal 0-5V, 0-10V, 4-20 mA
Start/Stop, change direction signal Passive switch signal, such as: foot pedal switch, Active switch signal: 5V-24V
Communication interface RS232 (standard), Hygiaflex-N232 Communication protocol; RS485 (optional), MODBUS protocol (RTU mode)
Power supply AC 220V ±10% 50Hz/60Hz (standard) AC 110V ±10% 50Hz/60Hz (optional)
Drive dimension (LxWxH) 230x152x198 (LxWxH)
Power consumption < 80W
Drive Weight 5.02kg
Condition temperature 0-40ºC
Relative humidity < 80%

Product Dimension Drawing (mm)

Product related

HF-EasyPump (Pressure Adjustable)

HF-EasyPump
(Fixed Pressure)

HF-EasyPump PPS (Pressure Adjustable)

HF-EasyPump PPS (Fixed Pressure)